导读: C CB 003, Venous Pressure 静脉压 C CB 003 000 CASE ERROR Transient error General description of conditions for occurring: - Internal software error 一般为软件错误,试着重新开始。 C CB 006, Blood Pump Control 血泵控制 C CB 006 000 CASE ERRO

C CB 003, Venous Pressure 静脉压

C CB 003 000

CASE ERROR

Transient error

General description of conditions for occurring:

- Internal software error 一般为软件错误,试着重新开始。

C CB 006, Blood Pump Control 血泵控制

C CB 006 000 CASE ERROR

Transient error General description of conditions for occurring: - Internal software error 一般为软件错误,试着重新开始。 C CB 006 001 ERROR DURING REGISTER LOAD/READ Unconditional error, blood pump not available General description of conditions for occurring: - The blood pump timer is not working properly. 血泵计时器工作不正常。 Technical description of conditions for occurring: - The timer 9513 has returned error message(s). 计时器9513返回错误信息。 Some possible explanations why the error has occurred: - The timer IC 34 on the control driver board is broken. 在控制驱动板上的IC34可能有问题。 C CB 006 002 ERROR DURING SET DISCRETE PORT Unconditional error, blood pump not available General description of conditions for occurring: - Internal software error 一般为软件错误,试着重新开始。

C CB 007, Art And Ven Clamp Control 动脉夹及静脉夹控制

C CB 007 000 CASE ERROR

Transient error General description of conditions for occurring: - Internal software error 一般为软件错误,试着重新开始。

C CB 008, Blood Volume 血液总量

C CB 008 000 CASE ERROR

Transient error General description of conditions for occurring: - Internal software error 一般为软件错误,试着重新开始。 C CB 008 001 ERROR DURING REGISTER LOAD/READ

LOAD/READ信号注册中出现错误。

Unconditional error, blood pump not available General description of conditions for occurring:

- The blood pump timer is not working properly. 血泵计时器工作不正确。 Technical description of conditions for occurring: - The timer 9513 has returned error message(s). 计时器9513返回错误信息。 Some possible explanations why the error has occurred: - The timer IC 34 on the control driver board is broken. 控制驱动板上的计时器IC34可能坏了。 C CB 008 002 BLOOD PUMP TACHO ERROR血泵上的TACHO信号错误。 Unconditional error, blood pump not available General description of conditions for occurring: - The blood pump tacho count-up value is incorrect 血泵TACHO信号累计量不正确。 Technical description of conditions for occurring: - Tacho count up value is out of limit (<0 or >20000) TACHO信号累计量超出范围(<0 或>20000) Some possible explanations why the error has occurred: - The timer IC 34 on the control driver board is broken 控制驱动板上的计时器IC34可能坏了。

C CB 009, Blood Flow血流量

C CB 009 000 CASE ERROR

Transient error General description of conditions for occurring: - Internal software error 一般为软件错误,试着重新开始。 C CB 009 001 ERROR DURING REGISTER LOAD/READ Unconditional error, blood pump not available General description of conditions for occurring: - The blood pump timer is not working properly. 血泵计时器工作不正常。 Technical description of conditions for occurring: - The timer 9513 has returned error message(s). 计时器9513返回错误信息。 Some possible explanations why the error has occurred: - The timer IC 34 on the control driver board is broken. 控制驱动板上的计时器IC34可能坏了。 C CB 009 002 DELTA TIME ERROR Unconditional error, blood pump not available General description of conditions for occurring: - Timing error. Technical description of conditions for occurring: - The time between the two last blood flow measurements was out of range. Some possible explanations why the error has occurred: - Internal software error. 一般为软件错误,试着重新开始。 - The timer IC 34 on the control driver board is broken. 控制驱动板上的计时器IC34可能坏了。 C CB 009 003 NUMBER OF TACHO PULSES ERROR Unconditional error, blood pump not available General description of conditions for occurring:

The number of counted blood pump tacho pulses is incorrect.

Technical description of conditions for occurring:

Either 0 or more than 10000 tacho pulses have been counted since the last time the blood flow was measured

Some possible explanations why the error has occurred: - No connection between tacho and BM I/O board 信息TACHO与BM I/O板之间连接有问题。

- Blood pump error (blood pump stopped)

血泵有故障。

- The timer IC 34 on the control driver board is broken.

控制驱动板上的计时器IC34可能坏了。 C CB 009 004 MEASURED BLOOD FLOW OUT OF LIMITS 检测时血流量超出范围。

Unconditional error, blood pump not available

General description of conditions for occurring: - The measured blood flow has differed too much from the blood flow set

value.

Technical description of conditions for occurring: - BRI(BI_BloodFlActSet) - BRI(BI_BloodFl) < -600

or

BRI(BI_BloodFlActSet) - BRI(BI_BloodFl) > 600

Some possible explanations why the error has occurred: - The blood pump unit is broken血泵传动轴有问题。 - The BM I/O board is broken (Q1-Q4 and components) BM I/O(Q1-Q4板有故障。和其他元器件。)

- The timer IC 34 on the control driver board is broken.

控制驱动板上的计时器IC34可能坏了。

C CB 010, Heparin Pump 肝素泵

C CB 010 000 CASE ERROR

Transient error General description of conditions for occurring: - Internal software error 一般为软件错误,试着重新开始。 C CB 010 001 ERROR DURING REGISTER LOAD/READ

LOAD/READ信号注册中出现错误。

Conditional error, confirm by zero-set heparin flow General description of conditions for occurring: - The heparin pump timer is not working properly 肝素泵计时间工作不正确。 Technical description of conditions for occurring: - The timer 9513 has returned error …… 此处隐藏:27985字,全部文档内容请下载后查看。喜欢就下载吧 ……

金宝AK95故障代码中文版.doc 将本文的Word文档下载到电脑,方便复制、编辑、收藏和打印
本文链接:https://www.jiaowen.net/wenku/1695363.html(转载请注明文章来源)
Copyright © 2020-2025 教文网 版权所有
声明 :本网站尊重并保护知识产权,根据《信息网络传播权保护条例》,如果我们转载的作品侵犯了您的权利,请在一个月内通知我们,我们会及时删除。
客服QQ:78024566 邮箱:78024566@qq.com
苏ICP备19068818号-2
Top
× 游客快捷下载通道(下载后可以自由复制和排版)
VIP包月下载
特价:29 元/月 原价:99元
低至 0.3 元/份 每月下载150
全站内容免费自由复制
VIP包月下载
特价:29 元/月 原价:99元
低至 0.3 元/份 每月下载150
全站内容免费自由复制
注:下载文档有可能出现无法下载或内容有问题,请联系客服协助您处理。
× 常见问题(客服时间:周一到周五 9:30-18:00)